OM-DAQ-USB-2401
Multiple Channel USB Data
Acquisition Module
For latest product manuals:
omegamanual.info
Shop online at
omega.com
®
Users Guide
®
Przeglądanie stron 0
1 2 3 4 5 6 ... 55 56

Podsumowanie treści

Strona 1 - User’s Guide

OM-DAQ-USB-2401Multiple Channel USB DataAcquisition Modulee-mail: [email protected] latest product manuals:omegamanual.infoShop online atomega.com®Use

Strona 2

Section 2 - HardwareIt is important that you read this manual completely and follow all safetyprecautions before operating this instrument.2.1 Package

Strona 3 - CONTENTS

Section 3 - Hardware Setup3.1 OM-DAQ-USB-2401 System Components Figure 3-1. OM-DAQ-USB-2401 Hardware and DimensionsThe external power port is used in

Strona 4

3.2.1 USB-powered Hubs These hubs draw all power from the host USB connector’s power pins. Thepower is used for hub internal functions and for the hub

Strona 5

3.3 Connecting Your OM-DAQ-USB-2401 Acquisition System to PCWhen using a power adaptor with your DAQ system, be sureto supply power (from the adaptor

Strona 6

3.4 Connecting Various Hardware SetupsON-DAQ-USB-2401 data acquisition systems range from simple to complex. Oneexample of a simple system is that of

Strona 7 - Section 1 - Introduction

3.4.2 Connection to USB-powered HubIn this example, two OM-DAQ-USB-2401s are connected by cable to individualports of a single USB-powered hub. Since

Strona 8 - 1.3 Statement on CE Marking

Figure 3-5. Connections to Self-Powered Hub and to USB-Powered HubUSB port locations vary from PC to PC.3.5 OM-DAQ-USB-2401 MountingThe OM-DAQ-USB-24

Strona 9 - Introduction

3.5.1 OM-DAQ-USB-2401 Wall MountingIn order to mount the OM-DAQ-USB-2401 to the wall, you will need to:1. Remove the rear screws from the OM-DAQ-USB-2

Strona 10 - Section 2 - Hardware

Hardware Setup33-83.5.2 OM-DAQ-USB-2401 DIN Rail MountingIn order to mount the OM-DAQ-USB-2401 to a DIN-Rail, you will need to:1. Attach the DIN-rail

Strona 11 - Section 3 - Hardware Setup

4-1Software4Section 4 - Software4.1 Getting StartedThe following program files are included on the Omega DAQ Central UserSoftware CD supplied with you

Strona 12 - Hardware Setup

Servicing North America:U.S.A.: Omega Engineering, Inc., One Omega Drive, P.O. Box 4047ISO 9001 Certified Stamford, CT 06907-0047 USAToll Free: 1-800-

Strona 13 - (OPTIONAL)

Software44-24.2.2 Software SetupInsert the DAQ Central Software CD that was included with your DAQ unit intothe CD-ROM drive on your PC. Your system s

Strona 14

Software44-3Figure 4-3. Confirm Installation ScreenThe setup wizard now has all the information to complete the installation of thesoftware on your P

Strona 15

Software44-44.3 DAQ Central Software OperationBelow is the basic flow of operation when using the software:1. Highlight desired device.2. Open Device

Strona 16 - 3.5 OM-DAQ-USB-2401 Mounting

TOOLBAR BUTTONSStart Button Initiates data acquisition for thehighlighted device. Even if a StartTrigger is set, this must be pressedto Arm the devic

Strona 17

4.3.1 Main Control Window Pull-Down OperationsBelow are the operations that are available in the Pull-Down Menus of the DAQCentral Main Control Window

Strona 18

Figure 4-7. Tools MenuExport Data File – To convert data file from binary file to .csv fileFgure 4-8. Help MenuCheck for Updates To check for the la

Strona 19 - Section 4 - Software

Section 4.3.2 Device ConfigurationThe Device Configuration window contains the Channel, Triggers, and Datatabs. Details for these tabs are below. Wh

Strona 20 - Figure 4-1. Welcome Screen

Configuration Parameter SettingsChannel The input channel’s name (not editable)AN1D = analog channel 1, differential inputAN1H = analog channel 1, si

Strona 21 - Software

Figure 4-13. Trigger OptionsThe Triggers tab is used to set up triggers for starting and stopping dataacquisition. The Start Trigger and the Stop Tri

Strona 22

For all Start Trigger types, the device must be armed by pushing the Start Button on the main window.Data storage and display will not begin until the

Strona 23

iOM-DAQ-USB-2400Multiple Channel USB Data Acquisition ModuleTABLE OFCONTENTSPageSection 1 - Introduction ...

Strona 24 - Figure 4-6. Device Menu

Software44-12Figure 4-18. Main Window ScreenFigure 4-19. Gauges Screen Figure 4-20. Bars ScreenFigure 4-21. Digital Screen Figure 4-22. Table Scr

Strona 25 - Fgure 4-8. Help Menu

Figure 4-23. Waveform ScreenEach Data Display can be customized by clicking the Options Icon .Each Data Display's Options screen lets you choos

Strona 26

Software44-144.4 Power Line RejectionThe OM-DAQ-USB-2401 Series can take readings while making use of 50/60Hzline cycle rejection. While in the line

Strona 27 - Figure 4-11. Units Options

The converter is comprised of a 4th-order, delta-sigma (Δ∑) analog-to-digital(A/D) converter modulator followed by a programmable digital filter. A fl

Strona 28 - Figure 4-13. Trigger Options

In DAQ applications, thermocouples should not be connectedsingle-ended. Doing so can result in noise and false readings.This is especially true when a

Strona 29 - Figure 4-16. Data Tab

Resolution (Effective Number of Bits - ENOB, RMS) – The number of reliabledata bits that exist for a signal’s measurement. The greater the resolution,

Strona 30

4.11 Thermocouple MeasurementsThe OM-DAQ-USB-2401 provides effortless thermocouple (TC) measurements.The unit includes built-in cold-junction compensa

Strona 31 - Figure 4-23. Waveform Screen

Section 5 – Signal Management 5.1 Channel Control and ExpansionIn the OM-DAQ-USB-2401 system, the quantities and types of DAQ units useddetermine the

Strona 32 - 4.6 A/D Conversion

Scan Rate vs. Resolution2ResolutionScan Rate3,4(Scans (ENOB5Speed Designation Scan Time1(s) per Second - SPS) RMS) Very Slow, 50 / 60 Hz

Strona 33 - 4.7 Input Ranges

The following table provides general advice regarding the selection of ScanTime. The concepts are further illustrated by the figure, Examples of Under

Strona 34 - Analog Input Channels Diagram

iiOM-DAQ-USB-2400Multiple Channel USB Data Acquisition ModuleTABLE OFCONTENTSPage5.4 Triggering ...

Strona 35 - SCAN PERIOD

5.4 TriggeringTriggering controls an acquisition cycle. Once the system is armed, a trigger isrequired to collect the data.The user must determine the

Strona 36

Differential mode refers to a mode, or circuit set-up, in which a voltage ismeasured between two signal lines. The measured differential voltage is us

Strona 37 - Signal Management

Resolution: An analog-to-digital converter (ADC) converts an analog voltage toa digital number. The digital number represents the input voltage in dis

Strona 38

5.9 Input and Source ImpedanceAs illustrated below, the input impedance (Ri) of an analog-to-digital convertercombines with the transducer’s source im

Strona 39

Section 6 - CE Conformity6.1 OM-DAQ-USB-2401 Design for CE Conformity The OM-DAQ-USB-2401 has been designed to meet requirements as outlined inEuropea

Strona 40 - 5.5 Signal Modes

Figure 6-2 Ferrite CoreThe ferrite cores may be removed from the DAQ unit; however, in order toconform to CE standards, all of the ferrite cores must

Strona 41

Section 7 – Troubleshooting7.1 Basic ChecklistPower – Check USB cable connection. The green “PWR IND” LED should be liton the DAQ device. If using a p

Strona 42 - 5.8 Analog Filtering

Section 8 – Service & CalibrationYour Omega DAQ components have been built and factory calibrated to meet orexceed the specifications listed here

Strona 43 - 5.10 Crosstalk

Section 9 – SpecificationsGeneral Isolation: 500V from PCExternal Excitation Output Voltage: +12v DC Regulated, Maximum total current output 67mAPower

Strona 44 - (MIDPOINT OF CABLES)

Thermocouple Input Range: Type J: -18 to 1200°C (0 to 2192°F)Type K: -129 to 1372°C (-200 to 2502°F)Type T: -101 to 400°C (-150 to 752°F)Type E: -184

Strona 45 - Figure 6-2 Ferrite Core

iiiList of FiguresSection Figure Description ... PageSection 1 1-1 IEC Symbols ...

Strona 46 - Section 7 – Troubleshooting

Table 9-1. Thermocouple Accuracy (ºC) for OM-DAQ-USB-2401TC Type Temp. Typical Fast(°C) Slow-18 1.3 3.8J0 1.131200 0.9 2.5-129 1.8 5K0 1.4 3.91372 1.

Strona 47 - Service and Calibration

10.1 International Usage & CE Marking The OM-DAQ-USB-2401 multiple channel USB data acquisition module is CEmarked and certified for use in severa

Strona 50

NOTES:10-4

Strona 51

WARRANTY/DISCLAIMEROMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for aperiod of 13 months from date of

Strona 52

M5025/0112Where Do I Find Everything I Need forProcess Measurement and Control?OMEGA…Of Course!Shop online at omega.comSMTEMPERATURE䡺⻬Thermocouple, RT

Strona 53

ivList of FiguresSection Figure Description ... Page4-25 OM-DAQ-USB-2400 Block Diagram ...

Strona 54

Section 1 - Introduction Please read this manual completely before installing and operating your Omegadata acquisition system. It’s important to read

Strona 55 - RETURN REQUESTS/INQUIRIES

When using OM-DAQ-USB-2401 unit to acquire data,computer energy save modes can cause false data readings.Prior to using the unit, ensure your computer

Strona 56 - Shop online at omega.com

*At highest scan rate, 1000 samples/sec +/- 1% with onechannel on, +/- 5% with all channels on.The main items you will be using in your data acquisiti